Page1
Sampling Inlets for Hiden
Mass Spectrometers
Gas, Vapour and Liquid Sampling Systems
Hiden Gas Analysis systems are versatile. Inlet systems are user interchangeable, allowing for
transformation of your mass spectrometer to suit new or multiple application requirements.
Page2
Gas Sampling Inlets
For a wide range of sample flow rates, and sample pressure ranges
STANDARD QIC – QUARTZ INERT CAPILLARY
LOW PRESSURE STAINLESS STEEL CAPILLARY
X Sample pressure range 2 bar to 100 mbar
Operates to 200°C X Sample pressure range 250 to 25 mbar
X
X Capillary length: 1.8 m (standard) X Operates to 200°C
X Fast 300 ms sampling response X Capillary length: 0.9 m
X Sample consumption rates from 0.8 to 20 ml/min X All stainless steel assembly
X Swagelok 1/16” or 1/8” connection X Swagelok 1/16” or 1/8” connection
X Corrosion resistant option available X Interchangeable with regular capillary in standard
QIC systems
X Fast response 0.9 m capillary also available, with sample
consumption rates from 1.6 to 20 ml/min X Corrosion resistant option available
X Flexible construction
ULTRA-LOW FLOW QIC
HEATED CAPILLARY INLET
HT/HP GAS SAMPLER
X For evolved gas studies. For OEMS, online electrochemical
MS, where evolved gas flow is limited
X Sample pressure range 0.1 to 30 bar
X Sample consumption rate of 250 µl/min
X Connects to standard QIC Inlet
X Operates to 200°C
X Heated to 200°C
2 | Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems
Page3
VERY LOW PRESSURE SAMPLING INLET
MICRO-FLOW INLET X Sample pressure range 10 to 1 mbar
X Sampling rate 12 µl/min. 24 µl/min option X Heated to 120°C
X Single or multi-capillary inlet option X Capillary length: 0.3 m
X 8 – 16 s response. Specially engineered to maintain a X Swagelok 1/8” connection
good response X Rigid construction
1 1m meetrtree sseemi-fllexxibiblele h heaetaetde hdo hseo s2e0 02°0C0°C
Saample ffrroomm A ALDLD
Massss sspeecctrtormometerter
2--Sttage 5m33//hrr
SSccrroollll ppuummpp
(not-shown)
(not-shown)
MasMs sapsesc tsrpomecettreor mturebtoe pru mp 70 l/s
turbo pump 70 l/s
Key - Use PDF reader layer function for additional information
Note - Diagram is for illustrative purposes only and is not to any exact scale or proportion
Fig.1 Sectional diagram HPR20 System with special inlet for monitoring ALD gases and vapours
HPR20_System_for_ALD_Monitoring_with_isolation_valve_01a C
ANALYTICAL
SGL – HEATED LOW PRESSURE
INLET WITH ISOLATION VALVE
TRI-VALVE SAMPLING HEAD X For sampling ALD processes with options from 1-5 mbar
or 5-50 mbar
X Sample pressure range <10-4 mbar to 1000 mbar X Heated to 200°C
X Flexible connection to process chamber X Air actuated heated (200°C) isolation ‘ALD’ valve included,
X Automatic sample line switching not shown
Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems | 3
Page4
Multistream Gas Sampling Inlets
MSV MULTI-STREAM SELECTOR
X 4, 8 or 16-way available
X Heating option (160°C)
X Corrosion resistant option
X Sample flow rate : 50 ml/min
X Flow meter option
PROTEUS TWIN CAPILLARY INLET
X 20, 40, 80-way multi-port valve
X Typical application – measuring input and output of
X Heated to 120°C reaction vessel
X Standard flow > 100 ml/min X Heated to 160°C
X Low flow configuration – from 4 ml/min X Fast switching valve between capillaries - less than
X High flow configuration – to 10 l/min one second
X No crosstalk between streams X Two capillaries of either 0.9 m or 1.8 m
4 | Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems
Page5
Gas Sampling Inlets - Thermal Analysis
For furnace, glovebox, reactor and evolved gas analysis
QIC HT250 INLET
X Sample pressure range 2 bar to 100 mbar
X Heated to 250°C (HT250) / 450°C (HT450)
X
HOT-ZONE INLET Sample consumption rates from 0.8 to 20 ml/min
X Capillary length: 1.8 m
X For sampling from furnaces; interface from furnace to QIC X Swagelok 1/16” or 1/8” connection
X Includes heated filter assembly X Interchangeable with standard QIC Inlet
X Allows direct sample analysis to 1700°C X Corrosion resistant option available
X General design to suit most furnaces X Flexible and robust steel hose
QIC GLOVEBOX
X Heated to 200°C
X Provides low dead volume interface with glovebox
TGA-MS
QIC HEATED EXTENSION
X A range of interfaces are available for TGA-MS evolved
gas analysis X Available as upgrade to standard QIC systems
X Specially designed to suit designated TGA instrument X Extends length by 2 m (standard extension) or 4 m, 6 m
X Most manufacturers and models covered on request
Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems | 5
Page6
Gas Sampling Inlets for
Analysis of dissolved gases in liquids
DIRECT MEMBRANE INLET PROBE CUVETTE
X For immersion and/or insertion in liquid or sludge samples, X Cuvette style sample cell for biofuel research
soil core sampling for example X Removable cover for light sensitive research
ENZYME KINETICS PROBE
DENITRIFICATION PROBE
X Probe sampler for the study of the chemical reactions that
X Flow through probe, low flow design for denitrification studies are catalysed by enzymes
X Allows for water vapour removal using liquid nitrogen cold trap X All glass probe construction
6 | Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems
Page7
LARGE CIRCULAR MEMBRANE CELL
DEMS DIRECT INLET
X Flow through dissolved species probe inlet, with liquid flow
X Probe sampler for in situ measurement of electrochemical connections
products X Larger surface area for fast and sensitive measurements
X Can be positioned at the electrodes of Redoxme™ cells X Integrated thermocouple for on-line sample temperature
and/or users’ own cells measurement
DEMS
X A combined mass spectrometry inlet and electrochemical
cell for real time differential mass resolved determination of
gaseous or volatile electrochemical reactants, reaction
intermediates and products while scanning electrochemical
parameters FLOW-THROUGH DISSOLVED SPECIES PROBE
X Allows static or dynamic electrolyte flow X For analysis of sample with continuous flow, extracted sea
X Thin layer and dual layer cells or estuary water for example
Sampling Inlets for Hiden Mass Spectrometers Gas, Vapour and Liquid Sampling Systems | 7
Page8
Hiden’s quadrupole mass spectrometer systems address
a broad application range in:
GAS ANALYSIS
dynamic measurement of reaction gas streams
catalysis and thermal analysis
molecular beam studies
dissolved species probes Hiden Analytical Ltd.
fermentation, environmental 420 Europa Boulevard
and ecological studies Warrington WA5 7UN England
T +44 [0] 1925 445 225
F +44 [0] 1925 416 518
SURFACE ANALYSIS E info@hiden.co.uk
W www.HidenAnalytical.com
UHV TPD
SIMS - Quadrupole and ToF
end point detection in ion beam etch
elemental imaging – 3D mapping
Sales Offices:
We have sales offices situated around
the globe. Visit our website for further
PLASMA DIAGNOSTICS information.
plasma source characterisation
etch and deposition process reaction kinetic studies Hiden systems are supplied
analysis of neutral and radical species with at least one sampling
inlet, systems are user
upgradeable and can
accommodate a wide range
VACUUM ANALYSIS of inlet systems. To confirm
compatibility for specific
partial pressure measurement and control of process gases inlet systems for operation
reactive sputter process control with your system please
vacuum diagnostics contact Hiden Analytical or
vacuum coating process monitoring your local representative.
TDS 169/3