Page1
Data Acquisition System
Selection Guide
1586A Super-DAQ Precision Temperature Scanner and
2638A Hydra Series III Data Acquisition System/Digital Multimeter
1586A Super-DAQ 2638A Hydra Series III
Precision Temperature Scanner Data Acquisition System/Digital Multimeter
The Fluke Calibration 1586A and Fluke 2638A are two new data acquisition
systems that offer great value and performance. This selection guide will
help you understand the differences between the two, the strengths of each,
and which one will best meet your current and future needs.
Page2
1586A Super-DAQ
Precision Temperature Scanner
A temperature-optimized data logger that can “cross over” as a precision
thermometer for the lab
The 1586A should be considered by users focused on temperature data acquisition where accuracy
matters. It can be used with internal High-Capacity Modules, an external DAQ-STAQ Multiplexer, or
both. Configured with the internal High-Capacity Module, the 1586A has superior RTD and therm-
istor accuracy compared to the 2638A. Additional temperature-focused features that will appeal
to metrologists and technicians include sensor characterizations, a probe library, and the ability to
control Fluke Calibration dry-wells and baths in automated tests. For best measurement accuracy,
the 1586A accepts ITS-90, CVD, and polynomial temperature sensor characterizations. The 1586A
also measures and records lower level dc voltage and current.
The 1586A is ideal for data acquisition applications where temperature
accuracy is important, including:
• Chamber mapping
• Freeze-drying validation
• Heat treating test
• Materials testing
• Temperature sensor calibration
When configured with the external DAQ-STAQ Multiplexer, the 1586A provides convenient probe
connections and performance so good it is a great option for use as a reference thermometer read-
out for temperature calibration labs (accuracy improves to 0.29 °C for thermocouples and 0.005 °C
for RTDs). It’s a nice upgrade for users of thermometer readouts like the Black Stack or Chub E4 who
are interested in a more modern user interface and a more advanced data acquisition platform.
For more information on the 1586A visit www.flukecal.com/1586A.
2 Fluke Calibration Data Acquisition System Selection Guide
Page3
2638A Hydra Series III
Data Acquisition System/
Digital Multimeter
The perfect replacement for Fluke Hydra Data Loggers; more capable, too!
The 2638A is the successor to past Hydra models. It is a full-featured general purpose data acqui-
sition system and is a great fit for most data acquisition applications where very good measurement
accuracy is needed. Its thermocouple accuracy is better than competing data acquisition models,
and equal to that of the 1586A when using the Universal Input Module. The 2638A is expandable
to 60 channels per mainframe and over 1,000 channels with application software. Compared to
the 1586A Super-DAQ Precision Temperature Scanner, it features a faster scan rate of up to 45
channels/second, and covers higher ac and dc input voltages (up to 300 V vs. 50 V). The higher
voltage ranges and CAT II safety rating make it a robust solution for industrial factory and process
monitoring applications.
The 2638A is ideal for data acquisition applications within test and measure-
ment, industrial and manufacturing, including:
• Automated manufacturing test
• Component testing
• Oven profiling
• Burn-in testing
• Life cycle testing
• Process monitoring
• Materials testing
For more information on the 2638A visit www.flukecal.com/2638A.
Fluke Calibration Data Acquisition System Selection Guide 3
Page4
Features and specifications comparison
1586A with DAQ-STAQ 1586A with High-Capacity 2638A with Universal Input
Multiplexer Module Module
Maximum channel count 40 40 60
Maximum scan speed 10 (PRT) 10 (PRT) 16 (PRT)
(channel/second) 10 (TC) 10 (TC) 15 (TC)
10 (V dc) 10 (V dc) 45 (V dc)
Able to control Fluke Calibration Yes Yes No
dry-wells and baths in automated
tests
Current reversal (PRT only) Yes Yes No
Sensor characterizations ITS-90, CVD, Polynomial ITS-90, CVD, Polynomial Alpha, R0
Probe library Yes Yes No
TC accuracy 0.29 °C 0.62 °C 0.62 °C
(Internal CJC for type K at 0 °C)
RTD accuracy 0.005 °C 0.008 °C 0.038 °C
(4-wire, PT385, at 0 °C)
Thermistor accuracy 0.002 °C 0.002 °C 0.003 °C
(5 kΩ or 10 kΩ thermistor at 0 °C)
DC V input 50 V max, all functions and 50 V max, all functions and Front panel: 300 V max CAT II
ranges ranges Rear panel: 150 V max CAT II
DC V accuracy 0.0037 % + 0.0035 % 0.0037 % + 0.0035 % 0.0037 % + 0.0035 %
AC V (100 V) n/a n/a 300 V/150 V
AC V accuracy (60 Hz) n/a n/a 0.6 % + 0.08 %
DC I range 0 mA to 100 mA 0 mA to 100 mA 0 mA to 100 mA
DC I accuracy 0.015 % + 0.0011 % 0.015 % + 0.0011 % 0.015 % + 0.0011 %
Frequency range n/a n/a 3 Hz to 1 MHz
Safety/Conformity CE, CSA, IEC 61010 3rd ed. CE, CSA, IEC 61010 3rd ed. CE, CSA, IEC 61010 3rd ed.,
Category II
Fluke Calibration. Precision, performance, confidence.™
From other countries +1 (425) 446-5500 or
Fluke Calibration For more information call: Fax +1 (425) 446-5116
PO Box 9090, Everett, WA 98206 U.S.A. In the U.S.A. (877) 355-3225 or Web access: http://www.flukecal.com
Fluke Europe B.V. Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 ©2014 Fluke Calibration.
PO Box 1186, 5602 BD Specifications subject to change without notice.
Eindhoven, The Netherlands 2675 200 or Fax +31 (0) 40 2675 222 Printed in U.S.A. 1/2014 6001948A_EN
In Canada (800)-36-FLUKE or Pub-ID 12164-eng
Fax (905) 890-6866 Modification of this document is not permitted
without written permission from Fluke Calibration.